RT9 Series Multi-batch Automatic Dissolution System
A high-efficiency dissolution testing solution named RT900 Series Multi-batch Automatic Dissolution System performs 10 consecutive batch tests while requiring less human interaction. This system provides exact media distribution while delivering complete cleaning of vessels and sophisticated automatic Processes that ensure the reliability of dissolution experiments. The system enables USP 1/2 method processing to meet all regulatory specifications. The instrument contains a high-end autosampler that allows automated online filtering while providing unlimited data storage capabilities. Its precise temperature and speed control guarantee accurate and repeatable test results.
- Efficient 10-batch dissolution test
- High Precision Media Dispensing and Removal
- Thorough Cleaning for Dissolution Cup
- Advanced Fully Automated Experience
Key features
Automatically complete all 10 batches of dissolution testing tasks unattended.
High-end type autosampler that supports storage and online automatic replacement of secondary filters.
It can store at least 15 years of experimental methods and records without the risk of data overwriting.
Spraying without dead angle, with the appropriate cleaning fluid and cleaning times set to effectively reduce the risk of sample residue between multiple batches of experiments.
Applications
- General tablets
- Immediate-release tablets
- Enteric tablets
- Extended-release tablets
Technical specifications
- Dissolution Vessel Number
- 6
- Number of automatic experimental batches
- 10
- Setting Range
- 20-250 rpm (RPM)
- RPM Resolution
- 0.1 RPM
- Temp. Accuracy
- ≤ ± 0.5℃
- Temp. Resolution
- 0.1 ℃
- Level
- < 0.5°
- Vessel Verticality
- 90°± 0.5°
- Shaft Verticality
- 90°± 0.5°
- Centering Deviation
- < ± 2.0 mm
- Depth Positioning Deviation
- < ±1.0 mm
- Shaft Wobble
- < ± 1.0 mm
- Basket Wobble
- < ± 1.0 mm
- Temp. Control Rang
- Room Temp. ~ 45℃
- Degassing Effect
- Oxygen ≤ 3ppm
- Degassing Volume
- ≥50L
- Maximum Settable Sample Number
- 20
- Sampling Range
- 5 ~ 20 ml
- Sampling Precision
- ≤ ±1%
- —
- Meeting FDA21 CFR Part 11
More in this family
Ready to specify a Raytor system?
Tell us your method, dosage form and throughput. Our team will prepare a quotation.