FDA Identifies Data Integrity Failures in HPLC and UPLC Analyses

An FDA Warning Letter to Curia New York raises questions about access to chromatography software, unexplained injections and the traceability of results.

By the Advanced Editorial Team
Analyst reviewing chromatography data next to an HPLC system in a pharmaceutical laboratory
Illustrative image generated for this article; not a photograph of the inspected facility.

An inspection focused on the laboratory

On September 18, 2026, the U.S. Food and Drug Administration (FDA) issued a Warning Letter to Curia New York, Inc., citing deviations from Current Good Manufacturing Practice (CGMP) and data integrity concerns in its laboratory operations.

According to the letter, the inspection took place from January 28 to February 5, 2026, and documented deficiencies involving high-performance liquid chromatography (HPLC) and ultra-performance liquid chromatography (UPLC) analyses, as well as gaps in how electronic and paper records were controlled. The case is a reminder of how much depends on the controls placed around computerized systems used in pharmaceutical quality control.

Thousands of chromatographic injections without adequate explanation

The FDA states that laboratory analysts performed thousands of unexplained injections using Empower software over the previous 30 months, and more than 70 trial injections using TotalChrom in October 2025 alone.

The agency highlighted the use of the “Developer” role in Empower, rather than the “Operator” role, without appropriate documentation. In its response to the FDA, the company itself stated that nearly 20 percent of Empower trial injections had been acquired or altered in “Developer” mode.

Investigators cited specific examples on HPLC and UPLC equipment where trial injections could not be linked, through the available documentation, to a product, sample or analysis. Without complete justification, it becomes difficult to reconstruct what happened in the laboratory and to assess any potential impact on analytical results.

Access control and record integrity

The findings went beyond chromatography. Investigators observed an analyst using another analyst’s login credentials on a computerized system used in GMP work. Shared credentials undermine individual attribution and make it hard to establish who actually performed a given activity.

The FDA also reported finding original laboratory records in shred bins, including pH meter and analytical balance printouts and reviewer-signed data review checklists. The letter further questions a practice in which personnel reviewed photocopies of notebook pages instead of the original records.

In our reading, these observations show that data integrity depends as much on the consistent application of procedures and quality oversight as on the technology available.

The role of the audit trail in chromatography systems

The audit trail is an important element of data integrity in regulated laboratories. In computerized systems, it records events related to the creation, modification or deletion of electronic records, helping to reconstruct the activities performed.

Having this feature, however, does not remove the need for complementary controls. Key practices include:

  • Access profiles that match each user’s responsibilities.
  • Individual, non-shared credentials.
  • Risk-based periodic review of audit trails.
  • Documentation and justification of trial injections and reprocessing.
  • Preservation of original data and associated metadata.
  • Analyst training and supervision.

Implications for pharmaceutical laboratories

The Warning Letter shows that regulatory compliance does not rest solely on the analytical capability of instruments or the features of software. Even sophisticated chromatography systems require defined procedures, appropriate user management and effective oversight.

For laboratories operating HPLC and UPLC equipment, the case is an opportunity to review the controls that cover the lifecycle of analytical data. That review should consider not only instrument performance but also procedures, documentation, user responsibilities and record review mechanisms.

Advanced Life Sciences' Perspective

For Advanced Life Sciences, data integrity is not a regulatory formality. It is a non-negotiable principle of science and pharmaceutical quality.

Investing in state-of-the-art instruments, sophisticated software and advanced analytical technologies is not enough if the data they produce cannot be fully traced, reconstructed and demonstrated.

An analytical result without adequate traceability loses its credibility, regardless of how precise the instrument is.

The case presented by the FDA highlights an issue that deserves attention across the industry: technology, however advanced, does not replace professional responsibility, operational discipline and a strong quality culture.

Data integrity should be built into the laboratory’s daily practice, from the first chromatographic injection to the final approval of results.

Regulatory compliance should be the outcome of reliable processes, not a concern that arises only when an inspection is near.

Advanced Life Sciences believes that trust in the pharmaceutical industry begins with trust in the data behind its decisions. Because in science, producing results is not enough. You must be able to prove they are reliable.